ULTRASONIC TESTING PART 2
DEAD ZONE :- THE TRANSDUCER IS NOT READY TO RECIEVE SIGNALS FROM NEAR SURFACE FLAWS AS IT IS STILL VIBRATING DUE TO INITIAL PULSE.
The dead zone can be verified with an International Institute of Welding (IIW) calibration block. With the time base calibrated to 50 mm, and the transducer on position , the extent of the dead zone can be inferred to be either less than or greater than 5 mm.
Fresnel Zone (Near Field) The sound beam radiated close to the transducer consists of a tapering near field or Fresnel zone beginning at the crystal surface, followed by a spreading far field or Fraunhofer zone.
Fraunhofer Zone (Far Field) Beyond the near field is the Fraunhofer zone . Here the beam diverges and is referred to as beam spread. Through the far field the sound behaves similar to a beam of light, in that the sound pressure disperses according to the Inverse Square Law (i.e., the intensity is inversely proportional to the square of the distance) and travels in straight lines.
FORMULA FOR CALCULATING NEAR FIELD:-
N= D^2/4lambda (λ)
N= NEAR FIELD
D= DIAMETER
LAMBDA:- WAVELENGTHS
Φ = DIVERGENCE ANGLE= sin -1(1.22LAMBDA/D)
